Introduction

Multi-shell acquisition, particularly the HCP-style three-shell protocol, is widely adopted for advanced diffusion MRI beyond DTI. This protocol samples b-values of 1,000, 2,000, and 3,000 s/mm², each with 90 directions.

However, this approach has two key limitations:

This document outlines recommended acquisition strategies, including (1) a practical two-shell protocol using the built-in DTI sequence and (2) an advanced 23-shell DSI protocol with 258 directions.


Recommendation 1: Multi-Shell Acquisition Using a Built-In DTI Sequence

The built-in DTI protocol supports multi-shell acquisition compatible with advanced methods such as GQI, QSDR, and RDI.

Step 1: b = 3000 s/mm² (60 Directions, Minimum TE)

Step 2: b = 1500 s/mm² (30 Directions, Same TE)

Final Checks

Ensure identical coverage and TE; adjust NEX if needed for SNR. Save the protocols as, for example:


Recommendation 2: 23-Shell DSI Acquisition (b-value = 0 to 4,000 across 258 Directions)

An advanced option is the 12-minute grid-258 acquisition, sampling 23 b-values from 0 to 4,000 s/mm² across 258 directions. Compared with HCP multi-shell acquisition, it reduces oversampling at low b-values and extends to higher b-values, improving sensitivity to restricted diffusion.

This grid scheme directly addresses the sampling and orientation-bias issues of HCP-style acquisitions.

With a multiband sequence (e.g., Siemens SMS or CMRR) at MB factor 4, this 2-mm, 258-direction acquisition completes in about 12 minutes.

Benefits of Grid Sampling:

Limitations:

Steps to Install the 12-Minute Grid Scheme on Siemens Prisma Scanners

Use the following files to set up the 12-minute 258-direction scan on Siemens Prisma. For faster acquisition, consider the 5-minute 101-direction scan.

Acquisition Notes

Sequence Configuration

  1. In the diffusion tab, set MDDW to Free mode.
  2. Place the b-table under C:\MedCom\MriCustomer\seq\. Rename any existing DiffusionVectors.txt, then copy the Grid-258 table to that folder as DiffusionVectors.txt.
  3. Set b-value1 = 0 and b-value2 = 4000.

Steps to Install the 12-Minute q-Space Scheme on Other Scanners

Convert the Grid-258 b-table to the required format for your scanner.

Acquisition Parameters

  1. Resolution: 2.0 mm isotropic (increase to 2.4 mm if SNR is insufficient).
  2. Matrix: 104 × 104.
  3. Slices: 72 (reduce if cerebellar coverage is not needed), no gap.
  4. Multiband acceleration: factor 3 or 4.
  5. Diffusion scheme: use bipolar diffusion encoding for eddy-current compensation. Monopolar acquisition with only post-processing eddy correction is not recommended for grid sampling.
  6. Use minimum TE and TR.
  7. Pixel bandwidth: approximately 1700 Hz/pixel.
  8. Phase encoding: anterior-to-posterior (A»P).
  9. Acquire an additional b0 image with reversed phase encoding (P»A) for distortion correction. Only the b0 image is needed for the reverse-phase acquisition.

Quality Check for Preliminary Results

  1. Verify that the brain contour remains visible in DWI images at b = 4000. If not, consider reducing the maximum b-value to 3000 to improve SNR.
  2. Generate SRC files from the diffusion MRI data. In DSI Studio, run Diffusion MRI Analysis → Step T1a: Quality Control and select the folder containing the SRC files. Review the Neighboring DWI Correlation values; low correlations may indicate motion artifacts or acquisition errors.

Conclusion

Following these guidelines provides practical multi-shell and grid-sampling options for advanced diffusion MRI. Proper quality control remains essential before model fitting, reconstruction, or interpretation.